NLZ E / NLZ
QWarning : HE1 - TS22798
A
If you continue to use a contra angle that was determined to be "NG" or "OIL" when the contra check function was performed, the contra heat prevention function will be activated and "HE1" will be displayed.
- Action
-
Perform maintenance of the handpiece and then check the handpiece with <Contra-Check> function.
QWarning : HE0 - TS22896
A
Overload or a risk of abnormal heating of the motor detected. <Motor overheat prevention function> *
- Action
-
Avoid overloading the handpiece with heavy cutting, etc. and wait for the warning to disappear.
* If you continue using the product, the motor will stop and error code “E8” will be displayed.
QError Code : E0 - TS22899
A
The motor shaft is locked.
- Action
-
Check if the handpiece is connected properly.
QError Code : E1 - TS22902
A
Over current in the circuit detected.
- Action
-
Avoid overloading the handpiece with continuous heavy cutting, etc.
Press the foot pedal again to remove the error.
QError Code : E2 - TS22906
A
Higher motor speed than rated value detected.
- Action
-
Avoid overloading the handpiece with continuous heavy cutting, etc.
Press the foot pedal again to remove the error.
QError Code : E3 - TS22909
A
The Motor drive IC (Integrated Circuit) chip on the board that controls the motor- has generated an error signal.
- Action
-
Avoid overloading the handpiece with continuous heavy cutting, etc.
Press the foot pedal again to remove the error.
QError Code : E4 - TS22912
A
High temperature of FET detected.
- Action
-
Allow the unit to cool down and try again.
QError Code : E5 - TS22915
A
Overvoltage input to the unit detected.
- Action
-
Check proper connection of AC adaptor and AC power cord.
QError Code : E6 - TS22918
A
Overvoltage of an LED light detected.
- Action
-
Check the LED light connection.
QError Code : E7 - TS22921
A
Detected a residual overvoltage when the motor was started.
- Action
-
Wait for approx. 5 sec., then press the foot pedal again.
If the error does not disappear, check if the motor and motor cord are connected properly.
QError Code : E8 - TS22924
A
A handpiece heat generation error was detected.
- Action
-
Perform maintenance of the handpiece, and then check the handpiece by the Contra- Check.
A heat generation error of the motor was detected.
- Action
-
Leave the motor until it cools down.
QError Code : E9 - TS22928
A
The motor does not start up.
- Action
-
Check the motor connection.
QError Code : EC - TS22931
A
Data cannot be saved to EPROM (Memory).
- Action
-
Turn the power OFF and then ON again.
QError Code : ED - TS22932
A
Signals from the sensor cannot be read.
- Action
-
Check the connection of the motor to motor cord.
QError Code : EE - TS22947
A
An Increased handpiece heat generation error was detected.
- Action
-
Perform maintenance of the handpiece, and then check the handpiece by the Contra- Check.
QError Code : EF - TS22950
A
Under-voltage input to the unit detected.
- Action
-
Check proper connection of AC adaptor and AC power cord.
QThe LCD Display does not light. - TS22953
A
- Cause1
-
The power Switch is OFF.
- Action1
-
Turn ON the power.
- Cause2
-
The AC adapter or AC power cord is not connected correctly.
- Action2
-
Check the connection.
- Cause3
-
Internal Fuse is blown.
- Action3
-
Contact your Authorized NSK Dealer.
QThe motor does not run. - TS22960
A
- Cause1
-
The tubing, motor cord, AC adapter or AC power cord is not connected correctly.
- Action1
-
Check the connection.
- Cause2
-
The air pressure is not given, or not connected properly from the dental unit.
- Action2
-
Check the air pressure of the dental unit.
- Cause3
-
The LCD Display shows an error code.
- Action3
-
Refer to the error code. (Refer to “7-1 Error Code”)
QThe rotation speed of the motor does not rise. - TS22964
A
- Cause
-
The air pressure of the dental unit is lower than the “Upper limit of air pressure”.
- Action
-
Set the Upper limit of the air pressure below the air pressure of the dental unit.
(Refer to “3-8 Changing the settings for various functions”)
QBeeps continue to sound when turning ON the switch. - TS22967
A
- Cause
-
You are stepping on the foot pedal when turning on the power switch. (Safety function)
- Action
-
Do not step the foot pedal, and turn on the power switch.
QThe LED does not light. - TS22970
A
- Cause
-
End of life expectancy reached.
- Action
-
Contact your Authorized NSK Dealer.
QThe motor heats up abnormally during rotation. - TS22977
A
- Cause
-
Coolant air is not given, or not properly connected from the dental unit.
- Action
-
Check the air pressure on the side of the dental unit.
QWater leakage - TS22981
A
- Cause
-
The tubing, motor cord is not connected correctly.
- Action
-
Check the connection. If it is happening inside the control unit, contact your Authorized NSK Dealer.
QSettings when the power is turned ON are different from the previous ones when turning off the power. - TS22992
A
- Cause
-
The power is turned OFF using the main unit’s power switch.
- Action
-
Previous settings will not be retained when turning OFF the main unit’s power switch. Turn OFF using the control unit’s Power Key.
Varios 970
QNo / Poor vibration. - TS22530
A
The Front Panel does not light, even if the Power Switch is ON.
- Cause1
-
The Power Cord or the Jack is disconnected.
- Solution1
-
Correctly insert the Power Cord or the Jack.
- Cause2
-
The Fuse is burned out.
- Solution2
-
Contact dealer.*
The Tip does not generate vibration, in spite of depressing the Foot Control.
- Cause1
-
The Tip is not tightened firmly.
- Solution1
-
Tighten the Tip until the Tip Wrench clicks.
- Cause2
-
Worn Tip.
- Solution2
-
Replace the Tip.
- Cause3
-
Power has not been correctly adjusted for the Tip.
- Solution3
-
Adjust the power on the Power Guide or Tip case label. Do not exceed.
- Cause4
-
The Foot Control is disconnected.
- Solution4
-
Connect the Foot Control correctly.
- Cause5
-
Failure of vibrator in the handpiece.
- Solution5
-
Contact dealer.*
- Cause6
-
Failure of internal components of the Foot Control.
- Solution6
-
Contact dealer.*
* Repairs cannot be made by the customer.
QThe Tip is bent or broken. - TS22607
A
- Cause
-
Power has not been properly adjusted for the Tip.
- Solution
-
Adjust the power level the Power Guide or Tip case label. Do not exceed.
QThe Tip is flying away. - TS22610
A
- Cause
-
The Tip is not tightened firmly.
- Solution
-
Tighten the Tip until the Tip Wrench clicks.
QNoise from the handpiece. - TS22613
A
- Cause1
-
Power has not been properly adjusted for the Tip.
- Solution1
-
Adjust the power level on the Power Guide or Tip case label. Do not exceed.
- Cause2
-
The Tip is not tightened firmly.
- Solution2
-
Tighten the Tip until the Tip Wrench clicks.
- Cause3
-
Failure of vibration in the handpiece or the Control Unit.
- Solution3
-
Contact dealer.*
* Repairs cannot be made by the customer.
QThe handpiece is overheating. - TS22616
A
- Cause1
-
Power has not been properly adjusted for the Tip.
- Solution1
-
Adjust the power level on the Power Guide or Tip case label. Do not exceed.
- Cause2
-
The Tip is not tightened firmly.
- Solution2
-
Tighten the Tip until the Tip Wrench clicks.
- Cause3
-
Failure of vibration in the handpiece or the Control Unit.
- Solution3
-
Contact dealer.*
* Repairs cannot be made by the customer.
QNo Irrigation supply and/or unstable Irrigation supply (Use of Bottle) - TS22619
A
The Irrigation Pump is running.
- Cause
-
The tube twisted.
- Solution
-
Straighten the twisted Irrigation Tube.
The Irrigation Pump is stopping.
- Cause
-
Time to replace Irrigation Pump. (Approx. 500hours after used.)
- Solution
-
Replace with new Irrigation Pump (Refer to 12. (4) Changing the Irrigation Pump).
QNo / Poor water. (Use of Tap Water) - TS22622
A
The water does not reach to the Control Unit.
- Solution
-
Check the water circuitry and supply to the Control Unit. Water pressure : 0.1-0.5MPa (1-5kgf/cm2)
Check to see if water reaches the Control Unit.
- Cause1
-
The Water Adjustment Knob is closed.
- Solution1
-
Turn the Water Adjustment Knob and adjust to the appropriate volume.
- Cause2
-
Disconnected Irrigation supply at low volume range. (less than 10ml/min.)
- Solution2
-
No problem. Turn the Water Adjustment Knob and increase the Irrigation volume.
- Cause3
-
The Water Filter is clogged.
- Solution3
-
Replace with new Water Filter (Refer to 12. (6) Changing Water Filter (Option) ).
QWater leakage. - TS22626
A
Water is leaking from the joint between the Irrigation Tube and the Irrigation Connector.
- Cause
-
The Irrigation Tube is not connected correctly.
- Solution
-
Firmly insert the Irrigation Tube into the Irrigation Connector inmost.
Water is leaking from the joint between the handpiece and the cord.
- Cause
-
O-Ring at the handpiece cord is worn or damaged.
- Solution
-
Replace with new O-Ring (Refer to 12. (3) Changing O-Ring •Handpiece Cord).
Water is leaking from the Control Unit.
- Cause
-
The water circuitry in the Control Unit is damaged.
- Solution
-
Contact dealer.*
* Repairs cannot be made by the customer.
QHandpiece LED does not illuminate. (Varios 970 LUX) - TS22630
A
Tip oscillates, but Handpiece LED turns on and off.
- Cause
-
The handpiece is not connected into the Handpiece Cord correctly.
- Solution
-
Firmly insert the handpiece into the Handpiece Cord inmost.
Tip oscillates, but Handpiece LED does not turn on.
- Cause
-
Disconnection in the Handpiece Cord, or failure in the Control Unit.
- Solution
-
Contact dealer.*
* Repairs cannot be made by the customer.
QStart Beeping - TS22633
A
Beeping while power on.
- Cause
-
Depress Foot Control.
- Solution
-
Release the Foot Control.
Beeping while stopping vibration of Tips.
- Cause
-
Abnormal heating of the Control Unit.
- Solution
-
Stop the operation and leave until Control Unit becomes cool.
Varios 370
QNo / Poor vibration - TS22640
A
With power on and no Power Indicator
- Cause
-
The AC Adaptor or the DC Plug is disconnected
- Solution
-
Correctly insert the AC Adaptor or the DC Plug
The Tip does not generate vibration, in spite of depressing the Foot Control
- Cause1
-
The Tip is not tightened firmly
- Solution1
-
Tighten the Tip until the Tip Wrench clicks
- Cause2
-
Worn Tip
- Solution2
-
Replace the Tip
- Cause3
-
Power has not been correctly adjusted for the Tip
- Solution3
-
According to Power Guide or Tip case label. Do not exceed recommendations.
- Cause4
-
The Foot Control is disconnected
- Solution4
-
Connect the Foot Control correctly
- Cause5
-
Failure of vibrator in the handpiece
- Solution5
-
Contact dealer*
- Cause6
-
Failure of internal components of the Foot Control
- Solution6
-
Contact dealer*
* Repairs cannot be made by the customer.
QThe Tip is bent or broken - TS22643
A
- Cause
-
Power has not been properly adjusted for the Tip
- Solution
-
According to Power Guide or Tip case label. Do not exceed recommendations.
QThe Tip is flying away - TS22646
A
- Cause
-
The Tip is not tightened firmly
- Solution
-
Tighten the Tip until the Tip Wrench clicks
QNoise from the handpiece - TS22651
A
- Cause1
-
Power has not been properly adjusted for the Tip
- Solution1
-
According to Power Guide or Tip case label. Do not exceed recommendations.
- Cause2
-
The Tip is not tightened firmly
- Solution2
-
Tighten the Tip until the Tip Wrench clicks
- Cause3
-
Failure of vibration in the handpiece or the Control Unit
- Solution3
-
Contact dealer*
* Repairs cannot be made by the customer.
QThe handpiece is overheating - TS22656
A
- Cause1
-
Power has not been properly adjusted for the Tip
- Solution1
-
According to Power Guide or Tip case label. Do not exceed recommendations.
- Cause2
-
The Tip is not tightened firmly
- Solution2
-
Tighten the Tip until the Tip Wrench clicks
- Cause3
-
Failure of vibration in the handpiece or the Control Unit
- Solution3
-
Contact dealer*
* Repairs cannot be made by the customer.
QNo / Poor water - TS22659
A
The water does not reach to the Control Unit
- Solution
-
Check the water circuitry and supply to the Control Unit. Water pressure : 0.1-0.5MPa (1-5kgf/cm2)
Check to see if water reaches the Control Unit
- Cause1
-
The Water Adjustment Knob is closed.
- Solution1
-
Turn the Water Volume Knob and adjust to the appropriate volume
- Cause2
-
Disconnected water supply at low volume range. (less than 10ml/min.)
- Solution2
-
Adjust the Water Volume Knob and increase the Irrigation volume
- Cause3
-
The Water Filter is clogged
- Solution3
-
Replace with new Water Filter (Refer to 11-3 Changing Water Filter (Option) )
QWater leakage - TS22662
A
Water is leaking from the joint between the handpiece and the cord
- Cause
-
O-Ring at the handpiece cord is worn or damaged
- Solution
-
Replace with new O-Ring (Refer to 11-2 Changing O-Ring Handpiece Cord)
QAttachment of the Control Unit Holder is loose - TS22665
A
- Cause
-
Holder worn
- Solution
-
Contact dealer*
* Repairs cannot be made by the customer.
QHandpiece LED does not illuminate.(Varios2 Lux) - TS22668
A
Tip oscillates, but Handpiece LED turns on and off
- Cause
-
The handpiece is not connected into the Handpiece Cord correctly
- Solution
-
Firmly insert the handpiece into the Handpiece Cord completely
QLoss of the power output without operation - TS22671
A
Power output is set 8 at G
- Cause
-
Protection Circuit is activated
- Solution
-
Power output will weaken automatically while continuous operation is over 10min at the setting of Maximum power at G mode. Releasing the foot from the Foot Control. Decrease the Power less than 5, once then increase the power again. (Refer to 7-5 Protection Circuit)
Surgic Pro2
QError Code: E0 – System error - TS18135
A
- Cause of error
-
Self-check error.
- Action
-
Request repair.
QError Code: E1 – Motor voltage/ current detection error - TS18034
A
- Cause of error
-
Extended use under heavy load.
Short circuit in the motor.
Motor cord (power line) failure.
- Action
-
Electrical contact may be insufficient.
Securely re-connect the motor cord.
If the error cannot be eliminated, request repair.
QError Code: E2 – Power supply error - TS18141
A
- Cause of error
-
Insufficient contact of the power cord.
Failure of the power supply circuit.
- Action
-
Securely re-connect the power cord.
If the error cannot be eliminated, request repair.
QError Code: E3 – Motor sensor error - TS18139
A
- Cause of error
-
Motor sensor failure.
Unplugged motor cord. Motor cord (signal line) failure. Ingress of water into the motor.
- Action
-
Electrical contact may be insufficient.
Securely re-connect the motor cord.
If the error cannot be eliminated, request repair.
(Make sure to attach a protection plug when cleaning in a washer-disinfector.)
QError Code: E4 – Control unit interior overheating error - TS18154
A
- Cause of error
-
Overheating by extended use under heavy load.
Use immediately after the control unit is placed at high temperature (such as inside a car under the blazing sun or in a locker exposed to direct sunlight).
- Action
-
Allow it to cool down before use.
To allow heat to be sufficiently radiated, the periphery of the control unit should be well-ventilated wherever possible.
If the error cannot be eliminated, request repair.
QError Code: E5 – Breaking error - TS18160
A
- Cause of error
-
Abnormal voltage generated in the start / stop switch circuit.
Failure in the start / stop switch circuit.
- Action
-
When rotation and stop are repeated in a short period of time, a circuit to suppress the sudden acceleration at the start of rotation may be activated. Wait for a few seconds before use.
If the error cannot be eliminated, request repair.
QError Code: E6 – Motor rotation failure error - TS18133
A
- Cause of error
-
Bur or drill attachment failure.
Handpiece failure.
Motor failure.
- Action
-
The chuck may be opened, or may not be sufficiently closed. Securely close the chuck.
If the error cannot be eliminated, request repair.
QError Code: E7 – Irrigation pump error - TS18165
A
- Cause of error
-
Incorrectly fitted irrigation tube caught on pump roller.
Irrigation pump failure.
- Action
-
Check the irrigation tube.
If the error cannot be eliminated, request repair.
QError Code: E8 – Bluetooth communication error - TS18163
A
- Cause of error
-
Failure of the Bluetooth communication circuit.
- Action
-
Request repair.
QError Code: E9 – Foot control error - TS18137
A
- Cause of error
-
Failure of the foot control sensor (Hall IC).
Self-check error.
- Action
-
Request repair.
QError Code: E10 – Motor LED error (Optic motor) - TS18159
A
- Cause of error
-
Adhesion of the saline solution to the motor LED.
Failure of LED circuit.
- Action
-
Remove the saline solution completely. If the error cannot be eliminated, request repair.
QError Code: E11 – Control panel error - TS18169
A
- Cause of error
-
Failure of the LCD display driver.
Failure of the electrostatic sheet.
- Action
-
Request repair.
QError Code: E13 – Osseo 100+ error - TS18156
A
- Cause of error
-
It is affected by electromagnetic interference from electronics Influence of electromagnetic interference waves emitted by electronic equipment.
Failure of Bluetooth communication circuit in Osseo 100+.
- Action
-
Keep the sufficient distance between the tip of Osseo 100+ and the electronic equipment. If the error cannot be eliminated, request repair.
QError Code: E14 – External terminal error - TS18167
A
- Cause of error
-
OS of external terminal is not supported.
Failure of the external terminal.
- Action
-
Check the supported OS in the application. Also, reinstall the application. If the error cannot be eliminated, request repair.
QScreen is blank. - TS23769
A
- Cause of problem1
-
Plug improperly inserted to power source
- Action1
-
Check to make sure the plug is in.
- Cause of problem2
-
Malfunction of power switch
- Action2
-
Request repair.
QThe maximum rotation speed and “–” are alternately displayed on the speed display. - TS23772
A
- Cause of problem1
-
Power turned on while pressing the foot control
- Action1
-
Remove foot from foot control.
- Cause of problem2
-
Foot control malfunction
- Action2
-
If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
- Cause of problem3
-
Control unit malfunction
- Action3
-
If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
QAfter the power is ON, “CONNECT” or “PUSH FC” is displayed and the screen is not switched. - TS23775
A
- Cause of problem
-
Foot control is not connected.
- Action
-
Keep pressing the speed control pedal of the foot control while “PUSH FC” is displayed on the LCD display. If the screen does not change after 5 minutes, press and hold the Cancel key. The normal screen is displayed.
Refer to “3.2 Installing foot control batteries” and replace the foot control batteries to check if it can be connected.
If it cannot be connected after replacing the batteries, Refer to “7.1.1 Pairing with foot control” for pairing.
If this does not resolve the issue, please contact an authorized NSK dealer.
QWill not rotate even if the foot control is pressed. - TS23776
A
- Cause of problem1
-
Foot control is not connected.
- Action1
-
Perform the following after confirming that nothing is blocking the signal between the foot control and control unit.
Turn the main unit power off and on again to connect to the control unit.
Refer to “3.2 Installing foot control batteries” and replace the foot control batteries with new ones.
Refer to “7.1.1 Pairing with foot control” and perform pairing.
If this does not resolve the issue, please contact an authorized NSK dealer.
- Cause of problem2
-
Foot control pedal and button response is poor due to not being used for a long period of time.
- Action2
-
Refer to “6.3 Foot control calibration” and calibrate the foot control.
- Cause of problem3
-
Malfunction of control unit or foot control
- Action3
-
If the problem is not solved by checking foot control batteries or foot control connection status, system repair is required.
Q“COM ERR FC” is displayed. - TS23781
A
- Cause of problem
-
Foot control not connected
- Action
-
Press the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and the foot control, wait for a while to check if the foot control is connected automatically. If it is not connected automatically, perform the following procedures.
Turn the control unit power off and on again to connect to the foot control.
Refer to “3.2 Installing foot control batteries” and replace the foot control batteries with new ones.
Refer to “7.1.1 Pairing with foot control” and perform pairing.
If this does not resolve the issue, please contact an authorized NSK dealer.
Q“COM ERR OS” is displayed. - TS23784
A
- Cause of problem
-
Osseo 100+ not connected
- Action
-
Press the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and Osseo 100+, wait for a while to check if Osseo 100+ is connected automatically. If it is not connected automatically, perform the following procedures.
• Turn off and on both this product and Osseo 100+ connected to this product.
• Charge the battery sufficiently with the Osseo 100+ charger and try the automatic connection again.
• Refer to the Osseo 100+ manual and try the pairing again.
If this does not resolve the issue, please contact an authorized NSK dealer.
Q“BATTERY FC” is displayed. - TS23787
A
- Cause of problem
-
Foot control battery is low
- Action
-
Depress the speed control pedal of the foot control or press the Cancel key to cancel the error.
Replace the batteries.
If this does not resolve the issue, please contact an authorized NSK dealer.
Q“BATTERY OS” is displayed. - TS23790
A
- Cause of problem
-
Osseo 100+ battery is low
- Action
-
Depress the speed control pedal of the foot control or press the Cancel key to cancel the error.
Charge the battery sufficiently with the Osseo 100+ charger and try the automatic connection again.
If this does not resolve the issue, please contact an authorized NSK dealer.